TruFluor pH
![]() Features
|
TruFluor pH
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Environmental Specifications | ||||
|---|---|---|---|---|
| Parameter | Units | Min | Typ | Max |
| Operating Temperature | °C | 4 | 45 | |
| Operating Pressure | Bar (psi) | 1(14.7) | 2(29.4) | |
| Storage Temperature | °C | 0 | 22 | 65 |
| Storage Relative Humidity (non-condensing) | % | 0 | 60 | 100 |
| Altitude | feet | 40,000 | ||
| Materials | USP Class VI and 316L electro-polishged SS | |||
| Performance Specifications | ||
|---|---|---|
| |
Min | Max |
| Sample Rate | Sample rate can be set so that samples are taken in increments from 1 second to 30 minutes | |
| Measuring Range (pH) | 5.5 | 8.5 |
| Specification Range (pH) | 5.5 | 7.5 |
| Relative Accuracy (pH) | +/- 0.1 pH units over +/- 0.5 pH range centered at 1-point standardization pH value | |
| Accuracy at 37.5 °C | Depends on accuracy of method and off-line analyzer used for 1-point standardization. | |
| Response Time (sec) | <90 (agitated, 90%) | |
| Drift | < 0.05 pH units over 21 days for a standard process sampling once every 5 seconds | |
| Calibration | Pre-calibrated (gamma-radiation-resistant chip). Conventional 2-point cal possible. 1-point standardization required. | |
| Gamma Radiation | Tested to 50 kGy (typical 25 - 40 kGy) | |
| Temperature | Accuracy | Precision |
| From 4 to 45°C | ± 0.25°C | ± 0.1°C |
| From 15 to 40°C | ± 0.15°C | ± 0.1°C |